A study of rate dependence in the energy measurement
-
Here the rate is referred as the number of incident
kaons (CK) and pions(Cpi ) per spill. Typically,
in the E949 experiment as shown in this
scatter plot, the Ck varies from 10 to 16 Million/spill,
while Cpi varies from 3 to 7 Million/spill. The ratio of Ck/Cpi
depends the experimental condition of the two separators in the beam line.
-
The cause of the rate dependence in the energy measurement
may be explained as the observed rate dependence of the gain factor
change in the Range Stack photo tubes and the accidental hits added in
the energy measurement.
To study the rate effect, we use Kp21 monitor
sample with some basic cuts (tpi-tk>2 ns, pr_rf > 0.02, |cos3d|
< 0.5, |chi_rm| < 3, ipiflg =0, trs < 45 ns and |Ptot
- 205| < 10 MeV/c). For a further study, we may apply some stringent
cuts (b4abm >1.2 MeV, b4trs < 2ns, cpitrs < 2, |b4str-tk| <
6ns, |tpi-trs| < 8ns, npe_cka>10, b4nhbm<6, cl_rsdedx > 0.02, itgqualt
<2,
nhitz > 4 and layv4 == tlay).
-
A
study of Kp2 energy peak as a function of rates shows: a) The Ck/Cpi
dependence can have a 0.25 MeV peak position variation, the same conclusion
can also be applied to Ck+Ck and Ck; b)
When Cpi less than 2 Million/spill (corresponding to Ck+Cpi
around
50% less than the normal rate), the peak shifts by 1 MeV. Further
study by comparing with the prediction from the UTC indicates this
one MeV shift is from the Range Stack Measurement. With the stringent cuts,
the results do not change ( See
this plot). All of which can be corrected
offline. Some
improvement is seen.
-
As expected, the
energy resolution varies with the rates: a) Tiny dependence on the
Ck/Cpi and Cpi ; b) Clear evidence on
the Ck+Cpi and Ck
dependence and
the magnitude is about 0.3 MeV. With the stringent cuts, the resolution
is improved but the dependence still persists (see
this plot). We might not be able to get
rid of this effect.